On this page ISO 15223-2 is the international standard that establishes a structured engineering and usability process for developing, selecting, and validating graphical symbols used on medical device labels, packaging, and accompanying documentation. While i

Illustrative imageIllustrative image

On this page

ISO 15223-2 is the international standard that establishes a structured engineering and usability process for developing, selecting, and validating graphical symbols used on medical device labels, packaging, and accompanying documentation. While its companion standard, ISO 15223-1, publishes the catalogue of recognized symbols, ISO 15223-2 defines the empirical testing required to prove that a symbol is readily understood by clinicians, lab staff, and patients before it enters clinical or commercial use.

This page describes standards, regulatory frameworks, and general quality management practices; it is not medical advice. Manufacturers should consult official standards texts and regional statutes when compiling labeling dossiers.

Definition and Scope of ISO 15223-2

Formally designated as BS ISO 15223-2:2010: Medical devices — Symbols to be used with medical device labels, labelling, and information to be supplied — Part 2: Symbol development, selection and validation, this standard was developed by ISO Technical Committee ISO/TC 210 in cooperation with national standards bodies such as the British Standards Institution (BSI Technical Committee CH/210/3). The primary driver behind ISO 15223-2 is overcoming language barriers in healthcare supply chains while mitigating usability risks associated with crowded packaging.

In global healthcare distribution, devices routinely cross international borders. When packaging must carry warnings, storage conditions, and handling instructions across multiple languages, physical label space rapidly diminishes. Manufacturers face challenging trade-offs: either enlarge packaging, increasing freight volume and packaging waste, or reduce font sizes, compromising legibility and clinical safety. Standardized graphical symbols provide an effective, language-independent solution, provided end users interpret them accurately and without delay.

The scope of ISO 15223-2 governs the procedural qualification of a symbol:

- Identifying the practical necessity for conveying critical information through a symbol rather than text.

- Formulating graphic concepts that adhere to ergonomic principles and international graphic standards.

- Assessing clinical risks arising from misinterpretation or confusion with existing icons.

- Conducting empirical comprehension and usability testing across representative user populations.

- Establishing objective acceptance criteria that determine whether a proposed symbol qualifies for inclusion in the ISO 15223-1 symbol catalogue or proprietary labeling files.

ISO 15223-2 does not govern electrical equipment safety signs (covered under IEC 60417), general public safety signs (ISO 7010), or radiation hazard pictograms. Instead, it serves as the dedicated benchmark for medical device and in vitro diagnostic (IVD) labeling symbology.

What is the Difference Between ISO 15223-1 and ISO 15223-2?

Understanding the distinction between Part 1 and Part 2 of the ISO 15223 series is essential for regulatory affairs professionals and packaging engineers. While both address medical device labeling, their operational roles are distinct.

ISO 15223-1 functions as an active reference library. It contains standardized, ready-to-use graphical symbols that manufacturers place directly on primary packaging, sterile barriers, and instructions for use (IFU). Familiar examples include Symbol 5.1.1 (the factory icon for Manufacturer), Symbol 5.1.4 (the hourglass for Use-by date), Symbol 5.2.1 (for Sterile), and Symbol 5.3.7 (for Temperature limit). When a manufacturer selects symbols directly from ISO 15223-1, regulatory authorities globally recognize them without requiring independent comprehension studies.

ISO 15223-2, by contrast, provides the empirical research rules and validation gates that must be satisfied before any new symbol can enter that library. Standardisation committees use Part 2 to evaluate proposals from industry working groups seeking to introduce new symbols into Part 1. Furthermore, medical device developers utilize Part 2 when designing proprietary symbols for novel therapeutic mechanisms or diagnostic indicators where no standardized icon currently exists in the medical device labeling symbols library.

The following table summarizes how the two parts interact across the device lifecycle:

The Five-Phase Symbol Development and Validation Process

ISO 15223-2 delineates a five-phase engineering and validation workflow designed to ensure that no ambiguous or misleading symbol enters medical distribution. The process requires cross-functional coordination between graphic designers, human factors engineers, clinical specialists, and regulatory managers.

Phase 1: Identifying the Need and Defining Scope (Clause 4)

The process initiates by evaluating whether an operational need truly warrants a new graphical symbol. Clause 4 requires developers to determine whether the proposed symbol has horizontal application across multiple device categories or whether it applies strictly to a restricted range of specialized instruments. Before commissioning artwork, developers must perform an audit of international databases, including ISO 7000 and IEC 60417, to verify an identical or conflicting concept does not already exist.

Phase 2: Classification of Usability and Safety Risk (Clause 6)

Clause 6 mandates that candidate symbols undergo formal risk classification based on the clinical consequences of user misinterpretation. In accordance with the ISO 14971 risk management framework, developers must evaluate failure modes:

- No to Low Safety Relevance: Misinterpreting the symbol might cause minor operational inconvenience or packaging confusion but presents negligible risk of patient harm or diagnostic delay (such as packaging recycling indicators or generic carton opening cues).

- Moderate to High Safety Relevance: Misinterpreting the symbol could directly result in patient injury, therapeutic failure, toxic exposure, or critical diagnostic errors (such as confusing single-use with reusable devices, misreading temperature extremes, or misinterpreting sterile barrier boundaries).

This risk classification directly dictates the statistical sample size and pass-fail criteria applied during empirical validation testing.

Phase 3: Concept Development and Graphic Ergonomics (Clause 7)

Once risks are categorized, graphic development proceeds under ergonomic principles codified in ISO 80416-1. Designers must craft visual metaphors that communicate abstract technical concepts clearly. Symbols must maintain visual fidelity when scaled down to minute label dimensions, such as a 5 mm square on an in vitro diagnostic reagent vial. Key graphic constraints include minimum stroke weights, uniform corner radii, high visual contrast, and the avoidance of intricate shading that could blur during high-speed printing.

Phase 4: Empirical Evaluation and Usability Testing (Clause 8)

Clause 8 defines the experimental protocols required to prove comprehension across real human users. Rather than relying on subjective opinions, ISO 15223-2 demands structured quantitative testing:

- Early Concept Screening (Clause 8.1): Multiple graphic variants are presented to representative user groups in an appropriateness ranking test, following ISO 9186-2 methods. Participants rank concepts based on intuitive clarity, allowing developers to eliminate weak visual metaphors early.

- Comprehensibility Testing (Clause 8.2): The leading candidate symbols undergo open-ended comprehensibility testing adhering to ISO 9186-1. Participants drawn from target user populations (such as practicing clinicians, laboratory scientists, or home-care patients) describe the symbol meaning without explanatory prompts. Responses are scored as correct, partially correct, incorrect, or opposite meaning (critical confusion).

- Memory and Retention Testing (Clause 8.3): If a symbol achieves borderline comprehension, developers test whether a brief instructional explanation enables users to retain and recall its meaning accurately over a subsequent evaluation cycle.

- Usability Walkthroughs (Clause 8.4): The final symbol is evaluated in simulated clinical environments under ISO 13485 quality management systems and IEC 62366-1 usability engineering standards to verify that clinicians notice and interpret the symbol correctly under actual task workflows and cognitive pressures.

Phase 5: Acceptance Criteria and Formal Adoption (Clause 9)

Clause 9 specifies the mathematical and qualitative thresholds that candidate symbols must achieve to be considered validated:

- Low-Risk Acceptance Thresholds: For symbols with negligible safety implications, a minimum comprehension rate of 67% in open-ended testing is generally acceptable, provided that no response indicates dangerous opposite actions.

- Moderate-to-High Risk Acceptance Thresholds: For symbols conveying safety-critical instructions, the standard demands higher comprehension levels (typically 80% or greater) and strictly enforces a 0% critical confusion tolerance. If even a small percentage of test subjects interpret a safety symbol as permitting an action that results in hazardous device misuse, the symbol fails validation and must be redesigned.

Regulatory Standing and the Presumption of Acceptable Risk

One of the most consequential legal and regulatory provisions in medical device governance is articulated in Clause 1 of ISO 15223-2. The standard explicitly states that if the symbol validation process detailed in Part 2 has been complied with, then the residual risks associated with the usability of that medical device symbol are presumed to be acceptable under ISO 14971 and IEC 62366, unless objective evidence demonstrates otherwise.

This presumption of acceptable risk provides medical device manufacturers with substantial regulatory certainty. When submitting technical documentation to notified bodies under the European Union Medical Device Regulation (EU MDR 2017/745) or the In Vitro Diagnostic Medical Device Regulation (EU IVDR 2017/746), demonstrating strict compliance with ISO 15223-2 establishes that the manufacturer has satisfied General Safety and Performance Requirement (GSPR) 23 regarding label clarity and usability. European guidance documents published by the Medical Device Coordination Group (MDCG) routinely affirm the primacy of harmonized international standards in establishing clinical usability.

In the United States, the Food and Drug Administration (FDA) implemented a landmark regulatory shift through its Final Rule published in the Federal Register (81 FR 38911, June 15, 2016), titled Use of Symbols in Labeling. This rule amended Title 21 of the Code of Federal Regulations, specifically 21 CFR § 801.15, authorizing manufacturers to use stand-alone symbols on medical device labeling without adjacent English explanatory text, provided two statutory conditions are satisfied:

- The symbol must be standardized in a standard recognized by the FDA through its consensus standards program (such as ISO 15223-1), or developed by a standards development organization (SDO) under an established validation process where the manufacturer demonstrates ordinary users understand its meaning.

- The medical device labeling must be accompanied by a comprehensive symbols glossary that explicitly defines every graphical symbol used on the packaging or device.

By complying with ISO 15223-2 validation methodologies, manufacturers generate empirical data required to justify stand-alone symbol implementation in FDA 510(k), De Novo, or Premarket Approval (PMA) regulatory submissions.

Empirical Testing Methodologies Compared

Conducting validation studies under ISO 15223-2 requires selecting the appropriate empirical testing methodology for each development milestone. The following table contrasts the primary testing modalities recognized under the standard, referencing normative testing frameworks from ISO 9186:

In addition to the summary above, the data table below compares the specific sample sizes, testing stages, and governing standards across the complete symbol development lifecycle.

Where ISO 15223-2 Matters for Labs, Clinics, and Supply Decisions

While ISO 15223-2 is an engineering and validation standard, its outcomes directly impact clinical workflow efficiency, diagnostic precision, and healthcare procurement decisions across hospitals and diagnostic laboratories.

Clinical Patient Safety

In high-pressure clinical environments, such as surgical suites, emergency departments, and intensive care units, healthcare workers make rapid decisions under intense cognitive load. Symbols validated under ISO 15223-2 ensure immediate, error-free recognition of critical safety parameters. For example, failing to recognize that a specialized vascular catheter is intended strictly for single-patient use can lead to hospital-acquired infections. The 0% critical confusion threshold enforced by ISO 15223-2 ensures that clinicians never mistake a single-use device for a reusable, autoclavable instrument.

Diagnostic Laboratory Operations

Clinical diagnostic laboratories handle thousands of in vitro diagnostic (IVD) test kits, microplates, and reagent cassettes daily. Reagents frequently carry precise environmental handling constraints, such as light sensitivity, humidity thresholds, and narrow storage temperature ranges (e.g., 2 °C to 8 °C). When IVD manufacturers validate symbols through ISO 15223-2, laboratory technicians can accurately sort and store inbound consumables within seconds of unpacking, preventing costly reagent degradation and false diagnostic readings.

Procurement and Supply Chain Standardization

Hospital materials managers and procurement officers manage inventory comprising tens of thousands of individual medical SKUs sourced from global manufacturers. Products that rely on standardized, ISO 15223-validated symbols reduce training requirements for supply warehouse personnel. Inventory handlers can verify sterility methods, batch numbers, and expiry dates at a glance, streamlining receiving audits, automated warehouse scanning, and first-expired, first-out (FEFO) stock rotation.

Common Misunderstandings and Implementation Pitfalls

Labeling compliance audits frequently reveal misconceptions regarding the scope and enforcement of ISO 15223-2. Regulatory teams must navigate several common pitfalls:

- Myth 1: Every manufacturer must validate every symbol they place on a box.

Reality: Manufacturers do not need to execute ISO 15223-2 validation studies for symbols that are already published in ISO 15223-1. Those symbols have already undergone international standardization and carry recognized consensus status. ISO 15223-2 is required only when proposing new symbols for standardization or when introducing proprietary, non-standardized graphics on commercial labeling.

- Myth 2: Symbol comprehension in one geographic region guarantees universal validity.

Reality: Graphical metaphors often carry unexpected cultural nuances. A visual cue that appears obvious to North American clinicians may carry alternate connotations or cause confusion among healthcare workers in East Asia, South America, or Europe. ISO 15223-2 strongly emphasizes testing cohorts that reflect the full geographical and cultural diversity of the intended market.

- Myth 3: Including an explanatory glossary excuses poor graphic design.

Reality: While 21 CFR 801.15 and EU MDR GSPR 23 require symbols glossaries to accompany labeled devices, a glossary cannot rescue an inherently confusing or hazardous graphic. In high-acuity clinical emergencies, practitioners look at the package, not the booklet. If a symbol fails empirical comprehensibility testing, it cannot be validated by relying on glossary text.

- Myth 4: Color coding can substitute for structural graphic clarity.

Reality: ISO 15223-2 symbols are designed as monochrome line graphics. Color is strictly controlled in international standardization because colored inks can fade under gamma radiation, ethylene oxide (EtO) sterilization, or prolonged light exposure. Furthermore, relying on color creates significant risks for color-blind clinical staff. Symbol validation must demonstrate that the graphic is fully legible and comprehensible in high-contrast black-and-white.

Frequently asked questions

What is the difference between ISO 15223-1 and ISO 15223-2?

ISO 15223-1 is the catalog of standardized medical device symbols used directly on product labels and packaging, such as manufacturer, expiry date, and batch code icons. ISO 15223-2 is the procedural standard that defines how new symbols must be developed, tested, and validated for user comprehension before they can be officially accepted or added to ISO 15223-1.

Who uses ISO 15223-2?

ISO 15223-2 is primarily used by international standardization committees (such as ISO/TC 210), medical device manufacturers designing proprietary labeling symbols for novel technologies, and human factors researchers conducting comprehension and usability studies for regulatory submission dossiers.

Why is ISO 15223-2 validation important?

Validation under ISO 15223-2 ensures that healthcare professionals, laboratory staff, and patients accurately understand the meaning of a symbol without verbal explanation. It eliminates language barriers, prevents packaging clutter, reduces clinical use errors, and establishes a formal presumption that usability risks under ISO 14971 and IEC 62366 are acceptable.

Do I need to comply with ISO 15223-2 to label my medical device?

In general, no. Medical device manufacturers labeling products with existing symbols from ISO 15223-1 do not need to perform ISO 15223-2 validation. However, if your device requires a novel or custom graphical symbol that is not present in ISO 15223-1, you must follow the ISO 15223-2 validation process to satisfy regulatory usability requirements under US FDA and EU MDR frameworks.

Sources

- BSI: BS ISO 15223-2:2010 | 30 Jun 2010 | BSI Knowledge, checked 2026-11-25

- Federal Register: Use of Symbols in Labeling, checked 2026-11-25

- Legal Information Institute: 21 CFR § 801.15 – Medical devices; prominence of required label statements; use of symbols in labeling, checked 2026-11-25

- European Commission: Guidance – MDCG endorsed documents and other guidance – Public Health, checked 2026-11-25

- World Health Organization: Health products policy and standards, checked 2026-11-25

Data: Medical Device Labeling Symbol Development and Validation Parameters

Structured comparison of symbol validation testing methodologies, normative references, participant cohort requirements, and statistical acceptance thresholds across ISO 15223-2, ISO 9186, ISO 14971, and 21 CFR 801.15.

Period covered: 2010/2026. Coverage: Global (ISO, US FDA, EU).

Sources

- BS ISO 15223-2:2010 | 30 Jun 2010 | BSI Knowledge (accessed 2026-11-25)

- Federal Register: Use of Symbols in Labeling (accessed 2026-11-25)

- Legal Information Institute: 21 CFR § 801.15 - Medical devices; prominence of required label statements; use of symbols in labeling (accessed 2026-11-25)

- European Commission: Guidance - MDCG endorsed documents and other guidance - Public Health (accessed 2026-11-25)

Compiled by CASRAI from the sources above. Free to reuse under CC BY 4.0 with attribution.